FDAJuly 13, 2026
Medline HUDSON RCI Breathing Circuits Recall
FDA published this notice on July 13, 2026 for United States.
This notice is kept at its address for anyone who saved the link. AreaZine no longer holds the agency's full record for it, so the page shows what AreaZine recorded and links to the agency's own page, the one to read.
What AreaZine recorded
- Agency
- FDA
- Place
- United States
- Date
- July 13, 2026
- Recall class
- Class II
- Reason given by the agency
- Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
- Distribution
- Worldwide distribution - US Nationwide and the country of Panama.
- Official source
- the notice at its source
Official warnings by state, with every type counted since 2008Search notices, places and declaration numbers