Boston Scientific StoneSmart Connect Console Recall Over Power Adapter Grounding Resistance
FDA published this notice on July 22, 2026 for Worldwide.
This notice is kept at its address for anyone who saved the link. AreaZine no longer holds the agency's full record for it, so the page shows what AreaZine recorded and links to the agency's own page, the one to read.
What AreaZine recorded
- Agency
- FDA
- Place
- Worldwide
- Date
- July 22, 2026
- Recall class
- Class II
- Reason given by the agency
- A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.
- Official source
- the notice at its source