Skip to main content
AREAZINENotices · US CA UK AU

Health Canada19 Nov 2026

VascuPuncture™ PICC High Performance Guidewire

Health Canada published a recall notice for VascuPuncture™ PICC High Performance Guidewire on 19 Nov 2026. It is one of 390 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.

Recall

The recall notice

Issued by
Health Canada
Issued
19 Nov 2026
Recall type
Health product recall
Companies
Heraeus Medical Components, LLC · 5030 Centerville Road, Saint Paul, Minnesota, United States, 55127
Official source
Health Canada: the notice at its source

The latest from Health CanadaEvery warning and recall in Canada

Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.

The notice as issued

Issue: Hydrophilic coated guidewire may swell inconsistently when exposed to saline, causing coating clumping and potential particle detachment. Released particles could enter the bloodstream and potentially cause embolic events. Users should immediately discontinue use of affected devices, quarantine inventory, and follow recall instructions for return. Additional information Details Original published date: 2026-11-19 Affected products: Affected products Lot or serial number Model or catalogue number VascuPuncture™ PICC High Performance Guidewire More than 10 numbers, contact manufacturer. 18070280NW-05MHP VascuPuncture™ PICC High Performance Guidewire More than 10 numbers, contact manufacturer. 18145280NW-05MHP Issue

Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.