Health Canada2 Oct 2026
Hexavue Focus Integrated System
Health Canada published a recall notice for Hexavue Focus Integrated System on 2 Oct 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 2 Oct 2026
- Recall type
- Health product recall
- Companies
- Steris Corporation · 2720 Gunter Park East, Montgomery, Alabama, United States, 36109
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: A software issue may result in unintended mismatch of patient information on manually printed images. During an active procedure, a user may manually select a different patient record from the “Patient Cases” menu and subsequently manually print images (“Manual Print Workflow”). In this specific Manual Print Workflow, the printed output may display patient identifying information from the selected patient record rather than the active procedure, including patient name, date of birth and medical record number (MRN).The image displayed is the image intended for the procedure; however, the printed image may contain incorrect patient identifiers. Auto-Print functionality is not impacted. Additional information Details Original published date: 2026-10-02 Affected products: Affected products Lot or serial number Model or catalogue number Hexavue Focus Integrated System SN: RL26028062 XEHFBU1 Hexavue Focus Integrated System SN: RL26063007 XEHFBU1 Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.