Giraffe™ OmniBed™ Carestation™

Health Canada published a recall notice for Giraffe™ OmniBed™ Carestation™ on 4 Jun 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.

Recall

The recall notice

Issued by
Health Canada
Issued
4 Jun 2026
Recall type
Health product recall
Companies
Datex-Ohmeda, Inc. · 3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188
Official source
Health Canada: the notice at its source

The latest from Health CanadaEvery warning and recall in Canada

Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.

The notice as issued

Issue: GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury. Additional information Details Original published date: 2026-06-04 Affected products: Affected products Lot or serial number Model or catalogue number Giraffe™ OmniBed™ Carestation™ More than 10 numbers, contact manufacturer. 2082844-001 Giraffe™ OmniBed™ More than 10 numbers, contact manufacturer. 2074920-001-XXXXXX Issue