The notice as issued
Issue: GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury. Additional information Details Original published date: 2026-06-04 Affected products: Affected products Lot or serial number Model or catalogue number Giraffe™ OmniBed™ Carestation™ More than 10 numbers, contact manufacturer. 2082844-001 Giraffe™ OmniBed™ More than 10 numbers, contact manufacturer. 2074920-001-XXXXXX Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.