Health Canada6 Oct 2026
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Health Canada published a recall notice for FreeO2 Automated Oxygen Therapy Device Adult & Pediatric on 6 Oct 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 6 Oct 2026
- Recall type
- Health product recall
- Companies
- Oxynov, Inc. · 725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39. Additional information Details Original published date: 2026-10-06 Affected products: Affected products Lot or serial number Model or catalogue number FreeO2 Automated Oxygen Therapy Device Adult & Pediatric Version HW 1.2 FO2-110-00 Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.