Skip to main content
AREAZINENotices · US CA UK AU

Health Canada6 Oct 2026

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric

Health Canada published a recall notice for FreeO2 Automated Oxygen Therapy Device Adult & Pediatric on 6 Oct 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.

Recall

The recall notice

Issued by
Health Canada
Issued
6 Oct 2026
Recall type
Health product recall
Companies
Oxynov, Inc. · 725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4
Official source
Health Canada: the notice at its source

The latest from Health CanadaEvery warning and recall in Canada

Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.

The notice as issued

Issue: A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39. Additional information Details Original published date: 2026-10-06 Affected products: Affected products Lot or serial number Model or catalogue number FreeO2 Automated Oxygen Therapy Device Adult & Pediatric Version HW 1.2 FO2-110-00 Issue

Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.