Skip to main content
AREAZINENotices · US CA UK AU

Health Canada24 Nov 2026

CARESITE Micro® Luer Access Device and Extension Sets

Health Canada published a recall notice for CARESITE Micro® Luer Access Device and Extension Sets on 24 Nov 2026. It is one of 385 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.

Recall

The recall notice

Issued by
Health Canada
Issued
24 Nov 2026
Recall type
Health product recall
Companies
B. Braun Medical Inc. · 824 12th Avenue, Bethlehem, Pennsylvania, United States, 18018
Official source
Health Canada: the notice at its source

The latest from Health CanadaEvery warning and recall in Canada

Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.

The notice as issued

Issue: B. Braun Medical Inc. (BBMI) is conducting a voluntary medical device correction for lots of CARESITE Micro® Luer Access Devices due to the piston remaining depressed following disconnection from the male luer. The depressed piston conidition may prevent valve closure and result in leakage. Additional information Details Original published date: 2026-11-24 Affected products: Affected products Lot or serial number Model or catalogue number CARESITE Micro® Smallbore Extension Set 0062060935 354269 CARESITE Micro® Smallbore Extension Set 0062057465 354269 CARESITE Micro® Smallbore Extension Set OOVL062352 354269 CARESITE Micro® Smallbore Extension Set 0061908600 354276 CARESITE Micro® Smallbore Extension Set 0061925959 354265 CARESITE Micro® Smallbore Extension Set 0061990229 354265 CARESITE Micro® Smallbore Extension Set 0062047387 354265 CARESITE Micro® Smallbore Extension Set 0062057592 354265

Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.

The record around this notice