Health Canada19 Aug 2026
SPECTRALIS® with Flex Module
Health Canada published a recall notice for SPECTRALIS® with Flex Module on 19 Aug 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 19 Aug 2026
- Recall type
- Health product recall
- Companies
- Heidelberg Engineering GmbH · Max-Jarecki-Strasse 8, Heidelberg, Germany, 69115
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose. As a result, the Flex stand may tip over. Users shall schedule a service appointment as soon as possible so that the castor fastenings can be inspected and corrected. Additional information Details Original published date: 2026-08-19 Affected products: Affected products Lot or serial number Model or catalogue number SPECTRALIS® with Flex Module Spec-21986 S2610 SPECTRALIS® with Flex Module Spec-24546 S3610 SPECTRALIS® with Flex Module Spec-23776 S3610 SPECTRALIS® with Flex Module Spec-13403 S3610 SPECTRALIS® with Flex Module Spec-15400 S3610 Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.