Health Canada28 Jul 2026
IQecg® Digital ECG
Health Canada published a recall notice for IQecg® Digital ECG on 28 Jul 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 28 Jul 2026
- Recall type
- Health product recall
- Companies
- Midmark Corporation60 Vista Drive, Versailles, Ohio, United States, 45380
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: Affected IQecg® plugin software may produce incorrect ECG waveforms and inaccurate heart-rate calculations. Until corrected software is installed, users should manually assess the patient's heart rate before initiating therapeutic intervention and use the temporary workaround if the issue occurs. Midmark will coordinate installation of the corrected software. Additional information Details Original published date: 2026-07-28 Affected products: Affected products Lot or serial number Model or catalogue number IQecg® Digital ECG SW v10-v10.1.0.17 4-000-0061 IQecg® Digital ECG SW v10-v10.1.0.17 4-000-0062 Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.