Health Canada23 Jun 2026
Hoffmann® II Carbon Connecting Rod
Health Canada published a recall notice for Hoffmann® II Carbon Connecting Rod on 23 Jun 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 23 Jun 2026
- Recall type
- Health product recall
- Companies
- Stryker GmbhBohnackerweg 1, Selzach, Solothurn, Switzerland, 2545
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: Subsequent to product complaints, Stryker has confirmed that cracking of the Hoffmann® II Carbon Connecting Rods may occur following cleaning and sterilization. Additional information Details Original published date: 2026-06-23 Affected products: Affected products Lot or serial number Model or catalogue number Hoffmann® II Carbon Connecting Rod All lots. 50796060 Hoffmann® II Carbon Connecting Rod All lots. 50796090 Hoffmann® II Carbon Connecting Rod All lots. 50796150 Hoffmann® II Carbon Connecting Rod All lots. 50796050 Hoffmann® II Carbon Connecting Rod All lots. 50796040 Hoffmann® II Carbon Connecting Rod All lots. 50796120 Hoffmann® II Carbon Connecting Rod All lots. 50796030 Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.