Health Canada23 Apr 2026
Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm
Health Canada published a recall notice for Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm on 23 Apr 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 23 Apr 2026
- Recall type
- Health product recall
- Companies
- Stryker Gmbh · Bohnackerweg 1, Selzach, Solothurn, Switzerland, 2545
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: Stryker confirmed that certain lots of Hoffman 3 Universal Pin Chucks contained inhomogeneous coating with out-of-specification coating thickness and color. This nonconformance was discovered via product complaint. Additional information Details Original published date: 2026-04-23 Affected products: Affected products Lot or serial number Model or catalogue number Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm AR4000 49229050 Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm AN4261 49229050 Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm AU0794 49229050 Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm AP3761 49229050 Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm 1544AB 49229050 Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.