Health Canada17 Jun 2026
HALYARD* PUREZERO* Ultraviolet* Nitrile Exam Gloves
Health Canada published a recall notice for HALYARD* PUREZERO* Ultraviolet* Nitrile Exam Gloves on 17 Jun 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.
Recall
The recall notice
- Issued by
- Health Canada
- Issued
- 17 Jun 2026
- Recall type
- Health product recall
- Companies
- O & M Halyard, Inc.1220 Old Alpharetta Rd, Alpharetta, Georgia, United States, 30005
- Official source
- Health Canada: the notice at its source
The latest from Health CanadaEvery warning and recall in Canada
Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.
The notice as issued
Issue: The issue surrounding this field action is related to not meeting the minimum breakthrough time of 30 minutes for n-Heptane required for European Type B performance per ISO 374-1:2016+A1:2018 and EN 16523-1:2015+A1:2018. No other product codes or lots of gloves are impacted in this field action. Additional information Details Original published date: 2026-06-17 Affected products: Affected products Lot or serial number Model or catalogue number HALYARD* PUREZERO* Ultraviolet* Nitrile Exam Gloves SJ504832X LFS511MD Issue
Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.