Skip to main content
AREAZINENotices · US CA UK AU

Health Canada26 Aug 2026

Cyanosure DermaV Laser System

Health Canada published a recall notice for Cyanosure DermaV Laser System on 26 Aug 2026. It is one of 311 notices of this kind from Health Canada since 2 Apr 2026 in AreaZine's record.

Recall

The recall notice

Issued by
Health Canada
Issued
26 Aug 2026
Recall type
Health product recall
Companies
Cynosure Lutronic Technology Corporation · 55, Samwon-Ro, Deogyang-Gu, Goyang-Si, Gyeonggi-Do, South Korea, 10550
Official source
Health Canada: the notice at its source

The latest from Health CanadaEvery warning and recall in Canada

Source: Health Canada: the notice at its source. AreaZine rebuilds this page from the agency's record; the agency's page is the one to act on. Contains information licensed under the Open Government Licence – Canada.

The notice as issued

Issue: Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn. Additional information Details Original published date: 2026-08-26 Affected products: Affected products Lot or serial number Model or catalogue number DermaV Not applicable. 1011600210 Issue

Source: Health Canada: the notice at its source. The text is the agency's, unedited. Contains information licensed under the Open Government Licence – Canada.

The record around this notice